ISO 15189 temperature checklist

October 18, 2026
3

What UKAS assessors look for on temperature, under ISO 15189:2022

A summary for lab quality managers. It isn't UKAS guidance: your assessor's view is the one that counts.

1. Know every unit

(ISO 15189:2022, 6.3 Facilities and environmental conditions; 6.6 Reagents and consumables)

  • List every temperature-dependent unit: fridges, freezers, incubators, water baths, the rooms where analysers and reagents sit, and sample transport.
  • Record each unit's acceptable range, taken from the manufacturer's instructions or the method.
  • Give each unit a unique ID that matches its records and its sensor.

2. Record and review

(6.3)

  • Monitor each unit at a defined frequency. Continuous monitoring covers nights and weekends.
  • Keep the records retrievable for your retention period.
  • Show that someone reviewed the records and signed them off, with the reviewer and the date.

3. Alarms and response

(6.3; 8.7 Nonconformities and corrective actions)

  • Set alarm limits for each unit, and document who is alerted, including out of hours.
  • Record every alarm: the time, who responded and what they did.
  • Treat an excursion as a nonconformity. Assess its effect on samples and reagents, and record the corrective action.
  • Test the alarms, and keep the test record.

4. Calibration and traceability

(6.5 Equipment calibration and metrological traceability)

  • Calibrate the monitoring sensors with traceability to national standards. A UKAS ISO 17025 certificate shows this.
  • Match each certificate to the sensor's serial number, and check that it's in date.
  • Track due dates, and record what you did when a sensor was found out of tolerance.

5. Measurement uncertainty

(7.3.4; UKAS LAB 48)

  • Know the uncertainty of each monitoring sensor. It's on the calibration certificate.
  • Document your decision rule. For example, set the alarm inside the limit by the uncertainty, so a +8 °C limit alarms earlier. Our guide works through an example.

6. Walk-in rooms and large stores

  • ISO 15189 doesn't name mapping. For a walk-in cold room, a mapping study is the usual evidence that the probe represents the whole space.

7. Blood storage

Have ready for the assessor:

  • the unit list with its ranges;
  • 12 months of records, with the review sign-offs;
  • the alarm log and responses;
  • the alarm tests;
  • calibration certificates matched to the sensors;
  • the decision rule.

Sources: