FAQs
Straight answers, by topic
The questions inspectors, assessors and buyers ask us most, answered in plain words. Each answer is sourced.
Monitoring
Alarms, records and sites
What happens when an alarm goes off at night?
The alert goes to the people you name for that unit, by phone and email. Each response is recorded with who acted and what they did.
Who gets the alarm when a freezer fails at 3am?
Whoever you name for that freezer, and the next person if they don't respond. Each alarm records who acknowledged it and what they did.
Can one system cover pathology, pharmacy and the mortuary?
Yes. Each department sees its own units, limits and alarm routes, and each inspector gets the records for their area. University Hospitals of Leicester monitors 350+ assets across three hospitals on one system.
Can you cover several depots?
Yes. Each site keeps its own limits and alarm routes, and head office sees every site on one record. One 25-site food distributor sees Over 95% optimal performance on JTF, many sites higher.
Can you monitor freezers across several buildings?
Yes. Wireless sensors report through gateways, so freezers in different buildings sit on one record, grouped by department or research group.
Do you monitor vehicles?
Yes, on the same platform as the store, so you can show an auditor or customer the temperature of a load from dispatch to delivery.
Can we see one blood unit's storage history for the MHRA?
Yes. Each fridge, freezer and incubator keeps its readings, alarms and actions, so you can show the conditions a unit was stored in and what happened if anything went out of range.
Will our RP see what happened during an excursion?
Yes: the readings, the alarm, who responded and what they did, on one record they can review and sign off.
Can you take over our existing system?
Yes. We service and support older wireless systems, and can keep yours running while we plan a move with you.
What does it cost?
It depends on how many units and sites you have. We quote after the check, once we know what you need.
Calibration
Who calibrates what, and to which standard
What can you calibrate?
Thermometers, temperature probes, data loggers and wireless temperature sensors. Our UKAS accreditation covers −90 °C to +135 °C for probes and sensors read with their own display or logger, and −35 °C to +40 °C for whole loggers and sensors, in our Tamworth lab or at your site.
Who calibrates the sensors?
We arrange calibration for every sensor we supply: traceable to national standards as standard, or UKAS accredited where you need it. Temperature sensors are calibrated in our own lab in Tamworth or at your site; for UKAS work it is our UKAS accredited calibration laboratory No. 10723, accredited to ISO/IEC 17025 for temperature for probes in a dry block from −90 °C to +135 °C, and for whole loggers and wireless sensors in an air chamber from −35 °C to +40 °C. Humidity, CO₂ and other parameters go to accredited partner laboratories. Each sensor has its own certificate.
Which lab calibrates which sensor?
Temperature: our own lab, traceable as standard or UKAS accredited in our UKAS accredited calibration laboratory No. 10723. Humidity, CO₂ and other parameters: an accredited partner laboratory, which issues its own certificate under its own UKAS accreditation number. We tell you which you'll get before we start. We carry out 10,000+ sensor calibrations a year.
Do you calibrate humidity and CO₂?
Yes, through accredited partner laboratories. Our own UKAS accreditation covers temperature only; partner laboratories issue their own certificates under their own UKAS accreditation number.
Is every calibration UKAS accredited?
No. Traceable calibration is our standard: it is traceable to national standards but not UKAS accredited. UKAS accredited calibration is there where you need it, for example for an ISO 15189 laboratory or where a customer or regulator asks for it. We tell you which you're getting before we start.
Can you calibrate on site?
Yes. On-site calibration is inside our UKAS accreditation, where the site suits the work; we agree that with you first.
Can you calibrate other makes of sensor?
Yes, any make of temperature probe, logger or sensor within our scope.
Why does the uncertainty matter?
The certificate states it. If your alarm limits ignore it, a reading near the limit could be out of range without the alarm going off. We can help you set limits that allow for it.
How do we know when calibration is due?
Due dates are tracked against each sensor and we contact you before they lapse.
Mapping
Temperature mapping, as a managed service
Is the mapping UKAS-accredited?
No. Our UKAS accreditation (UKAS accredited calibration laboratory No. 10723) covers temperature calibration. Mapping is carried out in line with GDP 3.2.1 and recognised guidance (WHO TRS 961 Annex 9 Supplement 8, ISPE), to the protocol we agree with you.
Can you map our warehouse?
Yes. Temperature mapping is a JTF managed service: we plan, run and report the study, then place the monitoring on its results. Our UKAS accreditation covers temperature calibration, not mapping.
Can you map our cold store?
Yes. Temperature mapping is a JTF managed service: we plan, run and report the study, then place the monitoring at the hot and cold spots it finds.
How long does a study take?
It depends on the size of the space and what your risk assessment asks for. We agree the duration in the protocol, before we start.
Do we need summer and winter studies?
If seasonal heat or cold could push the space out of range, usually yes. We'll tell you which applies to you on the call.
When do we need to map again?
GDP asks for re-mapping after significant changes, such as new racking, a new cooler or a changed door, or on the interval your risk assessment sets. We can book repeat studies with you.
Do we have to use JTF monitoring afterwards?
No. The report and probe positions are yours.
We're moving labs. Can you help?
Yes. We can map the new rooms as a managed service, place the monitoring on the results and move sensors with the equipment, so your records don't break during the move.
Validation
IQ, OQ, PQ and DQ, and who does which
Is the monitoring system validated for GDP?
In three parts. JTF qualifies the hardware at your site (IQ and OQ), Sematics validates the platform against its requirements and verifies each release according to its risk, and you qualify your configuration and use (URS, DQ and PQ) with our support, under a quality and technical agreement. See validation.
Do we have to repeat Sematics' testing of the platform?
No. Sematics validates the platform and verifies each release according to its risk. You assess Sematics, approve your reliance on that evidence for the functions you use, and qualify what Sematics can't: your requirements, your configuration, notification to your people, your hardware and your procedures. You stay accountable for your use of the system, as Annex 11 expects.
How does the platform meet EU GMP Annex 11?
Sematics validates the platform under a validation lifecycle aligned to EU GMP Annex 11, and gives you the evidence for your supplier assessment, applying the stricter July 2025 consultation draft where it asks more. Ask us for the current status of the validation documents under the quality and technical agreement. For each Annex 11 area, who is responsible and how Sematics approaches it, see Annex 11.
What is DQ?
Design and configuration qualification. Your configuration is designed from your URS; the platform then exports the complete configuration, and DQ compares that baseline item by item with the URS. It replaces guesswork about how the system is set up with a record you can show.
Is hardware covered by the platform validation?
No. Hardware is qualified separately: JTF carries out IQ and OQ of the sensors, gateways and probes at your site, with a calibration certificate for each sensor.
What happens when Sematics releases an update?
Each release is assessed and verified by Sematics according to its risk. Major releases are notified in advance with the release documentation and a configuration regression report run against your current configuration export. You assess each release under your change control and check the system after deployment.
Do configuration changes mean requalifying?
Not usually. Routine changes (users, contacts, names) are recorded in the audit trail. Changes within the qualified design (a limit or delay within your URS bounds, another escalation tier) are made with a reason and approved as your procedure requires. Only a change to the design, such as a new site or limits outside your URS, goes through change control with the affected DQ and PQ tests.
We're moving from another monitoring system. What changes?
An HQ addendum covers the history you carry over, alongside the usual URS, DQ and PQ. We plan the overlap so there's no gap in the record.
Support and service levels
Response times, renewals and older systems
What are the response times?
For a P1, an engineer replies within 4 working hours on Standard or 1 working hour on Enhanced, and attends on site within 3 working days or 24 hours on working days once we've established a visit is needed. P2 responses are within 2 working days. The full table is above, and the terms are in your agreement.
What does SLA mean here?
The service level agreement is the part of your agreement that says how quickly we respond and what we do, for each kind of issue. It is written in plain words and attached to your order.
How does renewal work?
Before the end date we send you a new order for the next term. Nothing renews without you agreeing it.
Is calibration included?
Yes, annual calibration of the sensors your agreement covers. Calibration is traceable as standard, or UKAS accredited where you need it: temperature in our UKAS accredited calibration laboratory No. 10723, other parameters through accredited partners.
What happens to our records if we leave?
We export your data and reports for you to keep.
What does it cost?
It depends on your sites, the sensors covered and the support level. We quote once we know what you need.
Do you come out to every fault?
No. Most faults are fixed remotely or with a simple swap your team can do with our help. An engineer comes when the fix physically needs one.
Who installs the system?
Our own engineers, from the same team that supports it afterwards.
Can you train new staff later?
Yes. We can run training again when your team changes.
We already have an older wireless system. Do we have to replace it all at once?
No. We can keep older wireless systems running and move department by department. See older system support.
Which systems do you support?
Existing legacy installs, including Hanwell-protocol sensors, System 5 and IceSpy-era systems. We no longer supply legacy sensors: replacements and new units are current Sematics sensors. The list is on legacy sensors we still support.
Do we have to replace our system?
Not straight away. We keep your existing install supported while we plan the upgrade with you; when a legacy sensor fails, it is replaced with a current Sematics sensor.
Can you move our historical records?
It depends on what your system can export. We check that with you before you commit to a move, and the old records stay available during the overlap.
Can you calibrate our existing sensors?
Yes. Temperature sensors in our own UKAS accredited calibration laboratory No. 10723 or on site, within our accredited range; humidity and other parameters through accredited partners.
Compliance
What the rules ask for on temperature
Does GDP require continuous monitoring?
The guidelines don't use the word. They do require alarms that alert you to excursions from pre-defined storage conditions (3.3), which a manual check can't do overnight or at a weekend.
How long should we keep temperature records?
GDP asks for backups of computerised data to be kept for at least five years (3.3.1), and for distribution records to be kept for at least five years.
Does the mapping have to be UKAS-accredited?
The guidelines ask for mapping under representative conditions; they don't ask for an accredited study. Calibration of the monitoring equipment must be traceable to national or international standards.
Does this apply to GP practices and pharmacies?
GDP is for wholesale distribution. Vaccine fridges in GP practices and pharmacies follow the Green Book, chapter 3, which our checklist also covers.
Does our hospital pharmacy fall under GDP?
If it holds a wholesale dealer's authorisation, yes, and the MHRA will expect mapping, monitoring and calibration to GDP. See MHRA GDP.
Do we still need to record the fridge temperature by hand?
The Green Book asks for readings recorded each working day. Keep the daily reading each working day, as the Green Book asks and CQC expects in GP practices; continuous monitoring adds the nights and weekends a manual check misses. Agree in writing which record is the original.
What happens when a vaccine fridge goes out of range?
The alarm goes to the person you name, the readings show how long and how far it went, and the record holds what they did, ready for the cold chain report to UKHSA or the manufacturer. See temperature excursions.
What will a UKAS assessor ask about temperature?
Usually: how limits were set, what happens when an alarm goes off out of hours, who acted, and whether the sensors are calibrated with traceable certificates. The ISO 15189 explainer goes through it clause by clause.
Does BRCGS require automatic temperature monitoring?
Not in every case. It accepts recording equipment with alarms or a documented system of manual checks. Auditors and retail customers increasingly expect continuous records, and a manual check can't catch an overnight failure.
Do care homes have to monitor fridges continuously?
The CQC asks for daily records and action when readings are out of range. Continuous monitoring keeps that record for you and catches the nights and weekends a daily check misses.
What happens when a medicine fridge goes out of range?
The person on shift gets the alarm, the readings show how far and how long, and the record holds what they did and the advice they took. See temperature excursions.
Sources
- EU Guidelines on Good Distribution Practice of medicinal products for human use (2013/C 343/01), applied in Great Britain by the MHRA
- UKHSA, The Green Book, chapter 3: Storage, distribution and disposal of vaccines
- ISO 15189:2022 Medical laboratories: requirements for quality and competence
- BRCGS Global Standard for Food Safety, Issue 9
- CQC: Storing medicines in care homes and adult social care
- UKAS schedule of accreditation 10723 (JTF Wireless Ltd, calibration)
- EudraLex Volume 4, Annex 11: Computerised Systems
20-minute compliance check
Your question not here?
In 20 minutes we look at how you prove temperature to your inspector or auditor, and where the gaps are. You get a short list of what to fix first. No product tour.