Temperature records and data integrity

How long to keep fridge records, whether a printout is enough, and what the MHRA's data integrity guidance means for electronic monitoring systems.

Reviewed by Jake Farrow, Managing DirectorLast reviewed 19 October 2026

This page explains what makes a temperature record trustworthy to an inspector: who made it, when, whether it has been changed, and whether you can still produce it years later. The one thing to remember is that where a system captures readings electronically, the MHRA treats the electronic data as the original record, so a filed printout is not enough on its own.

Why it matters

The record is the only proof you have

Nobody can go back and measure last March's fridge temperature. If a batch is questioned, a patient has a reaction, or an inspector asks how you know the cold chain held, the record is the whole of your answer. That is why inspectors treat a gap, an unsigned entry or an unexplained correction as seriously as a warm fridge: each one means the record can no longer be relied on.

The MHRA Inspectorate's data integrity blog for GDP gives everyday examples. Shared multi-user spreadsheets can make it impossible to show who made an entry, when, or whether it was overwritten. A wholesaler's proposal to report quality-critical events over WhatsApp was refused. An inspector could not trust ambient min/max records because the thermometer was uncalibrated and staff did not know how to use it. None of these involve bad intent; they involve records that cannot prove what they claim.

What the rules say

GDP, then the MHRA data integrity guidance

For wholesalers the legal duty comes from regulation 43 of the Human Medicines Regulations 2012, and the record-keeping detail from EU GDP. Chapter 4 says records are made at the time each operation is undertaken and kept for the period in national legislation, but at least five years. Section 3.3.1 says a computerised monitoring system must be validated or verified before use, its data protected against accidental or unauthorised change, checked for accessibility and backed up regularly to a separate, secure location.

EU GDP does not describe audit trails or what counts as the original. That detail is in the MHRA 'GXP' Data Integrity Guidance and Definitions (Revision 1, March 2018), which applies to GDP as well as GMP and uses ALCOA (it notes there is no difference in expectations from 'ALCOA+'). In the MHRA's own words (sections 3.10 and 6.1), data should be attributable to the person generating it, legible and permanent, contemporaneous, the original record (or a certified true copy) and accurate, and data governance should also keep it complete, consistent, enduring and available throughout its lifecycle. The MHRA describes it as a minimum expectation applied in proportion to risk, so a small site is not expected to check its data forensically, but it is expected to have assessed where its records could go wrong.

Raw data, printouts and audit trails

The MHRA defines raw data as the first capture of information. Where that capture is electronic and dynamic, 'paper copies cannot be considered as raw data', and keeping printed weekly reports in preference to the electronic data is not appropriate. You can keep only a static record if you can justify that it preserves the original, including metadata, audit trails and configuration settings, but the MHRA warns this is 'likely to be demanding in its administration'. Printouts remain useful for review and signing, as long as the electronic record is kept as the original.

A paper chart filled in by reading a display can itself be the original record. The MHRA asks for forms to be available where the work happens, so nobody jots readings on scrap paper and copies them later, and for transcriptions to be checked by a second person or a validated system. Scanning paper charts and destroying the originals is allowed if each scan is a verified true copy.

Where a system captures or stores raw data, the MHRA expects an audit trail showing who changed what, when and why, which users cannot switch off. It expects individual logins rather than one shared account, even if that means buying extra licences, and administrator rights held by as few people as possible. For a monitoring system, the entries that matter include alarm limit changes, sensor reassignments, acknowledgements and any edits to readings.

Hybrid, cloud-hosted and long-term records

Many sites run a hybrid: a logger plus a daily paper check. The MHRA accepts this if you document what makes up the whole data set and review and keep all of it, with extra review because hybrid systems are vulnerable to changes nobody can attribute. If your monitoring is cloud-hosted by a supplier, the MHRA says you remain responsible: you should know who owns the data, where it is physically held, and how you and the authorities get timely access to data, metadata and audit trails for the whole retention period, with that written into a technical agreement.

A nightly backup is not an archive. The MHRA distinguishes backup (a copy for disaster recovery) from archive (protected long-term retention) and expects records to stay readable, with their metadata, after a system is retired. Retention periods come from your sector: at least five years under EU GDP and in SPS records guidance, at least one year plus the stock's full storage history under Green Book chapter 3, and at least ten years in original format for HTA-licensed tissue establishments. For GMP batch documentation, API and investigational product records, see EU GMP and the Orange Guide.

Which source says what about temperature records
ControlMHRA data integrity 2018EU GDP 2013Green Book ch 3SPS records 2026
Sets a minimum retention period·✓✓✓
Records made at the time of the activity✓✓··
Electronic data is the original, not the printout✓···
Audit trail and individual logins expected✓···
Regular, checked backups✓✓··

A tick means the source states it in the items in this library; a blank means it does not address it there.

What good looks like

A record an inspector can trust

Write down, once, which record is the original for each unit, and how the paper and electronic records link. Give everyone their own login. Make entries at the time and sign them. Use ink: that is our reading of the MHRA's expectation that data is "legible and permanent" (section 6.1) and of the pencil records cited as a major deficiency in the 2016 GDP deficiency data; record a missed reading as missed, with the reason, and never fill it in afterwards. Correct mistakes so the original stays visible.

Review the records on a set routine and document the review the way the MHRA asks: a positive statement of whether issues were found, the date and the reviewer's signature, with a risk-based look at the audit trail. If you rely on a courier's or 3PL's temperature summary, assess their data integrity controls first, because a summary hides the detail.

  • Daily: entries made at the time, attributable to a named person.
  • At each review: positive statement, date, signature, audit trail checked.
  • Periodically, as your procedure sets: test that backups restore and archived data is still readable.
  • At any change of system or supplier: an archiving plan agreed before switch-off.

Worked example

A spike at 03:10 that looks like a faulty sensor

This is an illustrative example. At the Tuesday review, a wholesaler's quality lead sees that one cold room sensor logged a single reading of 14 °C at 03:10, with readings either side at 5 °C. The other two sensors in the room show nothing unusual. It looks like a sensor fault, and the temptation is to delete the point.

The MHRA says data can only be excluded with documented, valid scientific justification that it does not represent what was measured, and the excluded value must be kept with the data set. So she raises a deviation instead. She checks the sensor and its calibration, looks at the audit trail for any configuration change around 03:10, and compares the other sensors. She records her conclusion against the reading, keeps the value visible, and the Responsible Person signs off the review with a positive statement that one issue was found and how it was closed.

Where it's unclear

Where the sources disagree

Retention periods differ between EU GDP, the Green Book, WHO and the HTA (how long to keep records). The MHRA accepts controlled hybrid systems, while the WHO data integrity guideline says replacing them should be a priority with a documented CAPA plan (hybrid records). CQC expects GP practices to keep a daily manual record even with a logger, and the MHRA treats the logger's stored data as the raw data (logger versus daily reading).

Wherever the sources differ, assess the risk and record your reasoning.

Inspection findings

What MHRA inspectors have found in temperature records

The MHRA GDP deficiency data for 2016 includes a Responsible Person who admitted adding fabricated readings to fill gaps in temperature records, records made in pencil, unsigned temperature records and unapproved corrections, all cited as major deficiencies. The same data shows cold chain deviations that were recorded but never put through CAPA.

Gaps found at a CQC inspection do not automatically mean stock was badly stored. PHE's Q&A on CQC inspection says not to dispose of vaccines on that basis alone, but to contact the screening and immunisation team and gather evidence such as stock turnover and the fridge's service history. The responsibility and inspection page covers who should be reviewing records, and monitoring and alarms covers the instruments that produce them.

In summary

The short version

  • Decide and document which record is the original for each fridge, room or vehicle.
  • Keep electronic monitoring data as the raw data; treat printouts as copies.
  • Use individual logins, keep audit trails switched on and review them in proportion to risk.
  • Make entries at the time, signed, and record missed readings as missed.
  • Never delete a suspect reading; investigate it and record the conclusion against it.
  • Know your retention period, and keep backups and a separate archive that stays readable.
  • For cloud-hosted systems, get data access, location and exit terms in writing.

How long to keep records

Retention periods, by setting

Not contradictions but differences: each period applies to its own records. Where one unit holds products from more than one regime, keep its records for the longest period that applies.

SettingRecordHow longBasis
Wholesale (GDP), GB and Northern IrelandAll GDP documentation: procedures, records and data, including storage and transport temperature records. Backups of computerised system data are covered separately in chapter 3.3.1.At least 5 years, or longer if national legislation says soLaw (guidelines made binding by HMR reg 43(1))EU GDP Guidelines 2013/C 343/01, chapter 4.2 ('Documents should be retained for the period stated in national legislation but at least five years') and 3.3.1 (backups at least five years). Human Medicines Regulations 2012 reg 43(1) makes compliance a licence condition: in GB via reg C17, in NI via the Commission guidelines.Guidelines of 5 November 2013 on Good Distribution Practice of medicinal products for human use (2013/C 343/01)Compliance is mandatory for wholesale dealer's licence holders under HMR 2012 reg 43(1).
Wholesale (GDP), IrelandTransaction records for each medicinal product received or dispatched; all GDP documentation, including temperature records, under the EU GDP GuidelinesAt least 5 yearsLawS.I. No. 538/2007, Schedule 2, para 8(1) (transaction records kept available for inspection 'for a period of not less than five years') and para 13 (must comply with the Commission GDP guidelines, which set at least 5 years in chapter 4.2)S.I. No. 538/2007, Medicinal Products (Control of Wholesale Distribution) Regulations 2007The HPRA GDP guide (IA-G0046 v8, chapter 4) adds that superseded master copies of procedures should be kept for at least 5 years. The HPRA temperature guide (IA-G0011 v3) sets no retention period of its own. Neither HPRA guide gives a 15-year figure for GDP.
Manufacturing (GMP), UK and IrelandBatch documentation, plus critical raw data (e.g. validation or stability) that supports the marketing authorisationBatch documentation: 1 year after batch expiry or at least 5 years after QP certification, whichever is longer. Critical raw data supporting the MA: while the authorisation remains in force.GuidanceEU GMP Guide (EudraLex Volume 4), Part I, chapter 4, paragraphs 4.11 (batch documentation) and 4.12 (other documentation). Applied in GB through the MHRA's adoption of the EU GMP Guide.EudraLex Volume 4, EU GMP Guide Part I, Chapter 4: Documentation (in operation from 30 June 2011)Compliance is a condition of a manufacturing licence. For investigational medicinal products, batch documentation is kept at least 5 years after the last clinical trial using the batch ends. 4.11 notes longer periods may apply to specific products such as ATMPs (see the ATMP row: 30 years for traceability data).
Active substance (API) manufacture (GMP Part II), UK and IrelandBatch production, control and distribution records for APIs and intermediates, including records of critical storage conditions (Part II 10.10) and calibration records (5.32)At least 1 year after the batch's expiry date. For APIs with retest dates: at least 3 years after the batch is completely distributed.GuidanceEU GMP Guide (EudraLex Volume 4), Part II (ICH Q7), paragraph 6.13. Applied in GB through the MHRA's adoption of the EU GMP Guide.EudraLex Volume 4, EU GMP Part II: Basic Requirements for Active Substances used as Starting MaterialsPart II also requires controls against gaps in electronic data (5.43) and a back-up system where a breakdown could permanently lose records (5.48).
Investigational medicinal products (clinical trial supply), GB and Northern IrelandBatch documentation for each IMP batch, including the storage and shipping records that support its certification; product specification fileAt least 5 years after the completion or formal discontinuation of the last clinical trial in which the batch was usedLawGB: Commission Directive 2003/94/EC, Article 9(1), as referenced by the Medicines for Human Use (Clinical Trials) Regulations 2004, Schedule 7 (as amended by SI 2025/538, in force 28 April 2026). Northern Ireland: Commission Delegated Regulation (EU) 2017/1569, Article 8(3).Commission Directive 2003/94/EC, Article 9 (Documentation)IMP samples are kept for at least 2 years after the last trial ends (2017/1569, Article 11). The MHRA says separate GB guidelines for IMPs, reflecting the 2010 Annex 13, will be published.
Clinical trials (trial master file and essential records), UKThe trial master file and essential records, which typically include IMP shipment and site storage temperature recordsAt least 25 years after the trial ends, for trials applied for on or after 28 April 2026. Trials under the old rules: at least 5 years.LawMedicines for Human Use (Clinical Trials) Regulations 2004, regulation 31A, as amended by SI 2025/538 (in force 28 April 2026)Clinical trials for medicines: archiving and retention of clinical trial recordsAn RP or QP at a clinical-supply depot should check whether its temperature records form part of a sponsor's essential records; if so, the 25 years applies to them, not the 5-year IMP batch period above.
Advanced therapy medicinal products (ATMPs)Traceability data for each ATMP and its starting and raw materials, including all substances coming into contact with the cells or tissuesAt least 30 years after the product's expiry date, or longer if the marketing authorisation requiresLawRegulation (EC) No 1394/2007, Article 15 (traceability), retained in GB law; GMP 4.11 notes that ATMPs may need longer periods than other batch documentationRegulation (EC) No 1394/2007 on advanced therapy medicinal products, Article 15The 30 years covers traceability data. Whether storage temperature records must be kept as long depends on what they support; many ATMP sites keep them with the traceability data.
Blood establishments, UKRecords of donor testing and of activity data (donor and donation totals, hospital blood banks supplied, components produced, recalls, serious adverse events and reactions) and donor information and eligibility records; separately, records of serious adverse eventsAt least 15 yearsLawBlood Safety and Quality Regulations 2005 (SI 2005/50), reg 7(4) (records 'for a minimum period of 15 years') and reg 7(1)(g) (serious adverse events, at least 15 years)The Blood Safety and Quality Regulations 2005, regulation 7 (blood establishment requirements)The 15-year rule covers the record types listed in reg 7. It is not a general rule for all blood establishment documents, and the regulation does not name temperature logs.
Blood traceability, UK blood establishmentsLabel information for each unit and all other records needed for full traceability of blood and blood componentsNot less than 30 yearsLawBlood Safety and Quality Regulations 2005 (SI 2005/50), reg 8(2) and (3), read with Directive 2002/98/EC Article 14(3) and Directive 2005/61/EC Article 4 (traceability data at least 30 years)The Blood Safety and Quality Regulations 2005, regulation 8 (labelling and traceability)Directive 2002/98/EC Article 13 sets the 15-year period for other blood establishment records (https://eur-lex.europa.eu/legal-content/EN/TXT/?uri=CELEX:32002L0098).
Hospital blood banks, UKTraceability data (Part 6 of the Schedule, as it applies to the blood bank) from receipt of each unit; records of serious adverse eventsTraceability data: not less than 30 years. Serious adverse event records: at least 15 years.LawBlood Safety and Quality Regulations 2005 (SI 2005/50), reg 9(1)(e) and reg 9(1)(f)The Blood Safety and Quality Regulations 2005, regulation 9 (hospital blood banks)The regulation does not set a period for temperature charts. See the RCPath row for the 15-year recommendation on blood bank fridge and freezer charts.
Blood facilities (wards, theatres, clinics), UKTraceability data held by facilities where blood is delivered or transfused (Part 6, Section B of the Schedule)Not less than 30 yearsLawBSQR 2005 reg 12A(1)(a) and reg 12B(1)(a)The Blood Safety and Quality Regulations 2005The regulations do not name temperature charts for facilities; the 30 years covers traceability data.
Blood bank refrigerators and freezers (transfusion laboratories), UKRefrigerator and freezer temperature charts for blood stored for transfusion15 yearsGuidanceRCPath and IBMS, The retention and storage of pathological records and specimens, 6th edition (October 2025), section 6.1.5 ('These should be kept for 15 years') and section 4.20, which refers to the BSQR 2005 as the appropriate standardThe retention and storage of pathological records and specimens, 6th edition (G031, October 2025)The figure is RCPath's. The BSQR text itself sets 15 years for specific records (reg 7, reg 9(1)(f)) and does not name temperature charts. This is UK guidance; Irish law sets no period for fridge charts.
Blood establishments and hospital blood banks, IrelandBlood establishment records of donor testing, activity data and donor information; traceability records (blood establishments, to the point of delivery to a hospital); hospital blood bank traceability data from receipt; facilities where blood is delivered or transfused (Schedule 1 data)Blood establishment records: at least 15 years. Traceability data (blood establishments and hospital blood banks): not less than 30 years.LawS.I. No. 360/2005, reg 9(4) (15 years), reg 10(2) (30 years, blood establishments) and reg 11(e) (30 years, hospital blood banks); S.I. No. 547/2006, reg 10S.I. No. 360/2005, European Communities (Quality and Safety of Human Blood and Blood Components) Regulations 2005The regulations name the IMB, now the HPRA. From 7 August 2027 the EU SoHO Regulation (EU) 2024/1938 repeals Directives 2002/98/EC and 2004/23/EC; its Article 42(6) keeps traceability data for at least 30 years from distribution, disposal or export (https://eur-lex.europa.eu/legal-content/EN/TXT/?uri=CELEX:32024R1938).
Tissues and cells for human application (HTA licensed), UKAll records critical to tissue and cell safety and quality, including raw data such as original temperature monitoring records; separately, data needed for traceability from donor to recipientCritical records and raw data: at least 10 years after expiry, clinical use or disposal, kept in original format. Traceability data: at least 30 years after clinical use.GuidanceHTA Guide to Quality and Safety Assurance for Human Tissues and Cells for Patient Treatment (January 2021), glossary ('Raw data'), paragraph 76 (10 years) and paragraphs 95 and 175 (30 years). The 30-year rule comes from Directive 2004/23/EC Article 8, applied by the Human Tissue (Quality and Safety for Human Application) Regulations 2007, Schedule 2.HTA Guide to Quality and Safety Assurance for Human Tissues and Cells for Patient Treatment (January 2021)The guide is issued under HTA directions that licence holders must follow. The clock runs from expiry, clinical use or disposal, not from when the record was made. After 10 years, raw data may be kept in consolidated form for traceability.
Other HTA sectors (post mortem, research, anatomy, public display, organ donation), UKTraceability recordsNo fixed period except organ donation and transplantation (30 years from organ retrieval). Research and anatomy: at least until the tissue is disposed of (research also: used up or will never be used again). Post mortem: HTA points to its Code of Practice B and endorses the RCPath guidance.GuidanceHTA Record retention FAQs (updated 3 October 2025): 'With the exception of the Human Application and Organ Donation and Transplantation sectors, the HTA is not prescriptive about the length of time for which traceability records should be retained.'HTA, Record retention FAQsThe same FAQ says that for HTA-licensed NHS trusts, all records created under the Human Tissue Act 2004 are kept for 30 years, under the NHS records management code. The HTA sets no temperature-record period for these sectors.
Tissues and cells for human application, IrelandRecords needed for identification and full traceability of each tissue or cell donation and unit, including products that come into contact with them; donor records needed for full traceability; all records, including raw data such as storage temperature records, critical to the safety and quality of the tissues and cellsCritical records and raw data, including storage temperature records: at least 10 years after expiry, clinical use or disposal. Traceability data: not less than 30 years after clinical use.LawS.I. No. 158/2006 reg 14(4) and Annex 1.4.4; S.I. No. 598/2007 reg 21, Sch 1 Part D para 7 and Part E para 7.S.I. No. 158/2006, European Communities (Quality and Safety of Human Tissues and Cells) Regulations 2006; S.I. No. 598/2007, European Communities (Human Tissues and Cells Traceability Requirements, Notification of Serious Adverse Reactions and Events and Certain Technical Requirements) Regulations 2007S.I. No. 598/2007 requires storage temperatures to be monitored and recorded and kept for at least 10 years after expiry, clinical use or disposal. From 7 August 2027 the EU SoHO Regulation (EU) 2024/1938 applies.
Vaccine storage (GP practices, pharmacies, clinics), UKVaccine refrigerator temperature recordsAt least 1 year, and long enough to cover the full storage history of any product in the fridge. The Green Book says 5 years will generally cover this.GuidanceUKHSA, Immunisation against infectious disease (the Green Book), chapter 3: 'The records should be readily accessible, be retained for at least one year, and cover the full storage history of any products contained in the fridge... retaining records for five years will generally enable the full storage history of the vaccines be accounted for.'Storage, distribution and disposal of vaccines: the Green Book, chapter 3The Green Book gives the reason: manufacturer shelf lives can be up to four years or longer.
Pharmacy medicines storage (hospital and community), EnglandMedicine storage temperature records: fridge and ambient readings, excursions, monitoring system downloads and calibration records; vaccine temperature logs; vaccine refrigerator maintenance logsStorage temperature records and vaccine temperature logs: at least 5 years. Vaccine refrigerator maintenance logs: 11 years.GuidanceNHS Specialist Pharmacy Service, stock management, waste and recall records in pharmacy (citing the GDP guidelines and the Green Book); the 11 years for maintenance logs is SPS citing the NHS England Records Management Code of PracticeStock management, waste and recall records in pharmacySPS advice, not law. It is the main NHS pharmacy reference for these periods. The 11-year figure comes from the [NHS England Records Management Code of Practice](https://transform.england.nhs.uk/information-governance/guidance/records-management-code/), which SPS cites.
Pathology laboratories and mortuaries, UKRefrigerator and freezer temperature records; internal quality control records; equipment maintenance and service logsTemperature records: primary traces at least 2 months if regularly summarised; summarised or untranscribed records 5 years for reagents, short-term specimens, mortuary fridges and research or QC storage; lifetime of the equipment for long-term (biobank) freezers; as long as the oldest sample for forensic freezers. IQC records: at least 5 years. Maintenance and service logs: lifetime of the instrument plus at least 4 years.GuidanceRCPath and IBMS, The retention and storage of pathological records and specimens, 6th edition (October 2025), sections 4.17, 4.20, 4.21 and 4.22The retention and storage of pathological records and specimens, 6th edition (G031, October 2025)The 6th edition replaces the 5th (2015). Blood transfusion fridge and freezer charts are 15 years (see the blood bank row). Accreditation records: RCPath gives at least 8 years or 2 ISO 15189:2022 assessment cycles (ISO 15189 itself sets no retention period).
Care homes, EnglandMedicines fridge and storage temperature recordsNo period setGuidanceNICE SC1 (Managing medicines in care homes) and the CQC guidance on storing medicines in fridges in care homes both require temperature records but give no retention periodNICE SC1, Managing medicines in care homes: recommendationsCQC page checked: https://www.cqc.org.uk/guidance-regulation/adult-social-care/medicines-information-adult-social-care-services/storing-medicines-fridges-care-homes. Where a care home holds vaccines, the Green Book row applies.
Food (BRCGS Food Safety certified sites)Quality and food safety records, including temperature recordsA defined period that takes account of legal and customer requirements and the product's shelf life (including any consumer extension such as freezing, where the label allows it); at minimum the shelf life plus 12 months.GuidanceBRCGS Global Standard for Food Safety, Issue 9, clause 3.3.2 (paraphrased)BRCGS Food Safety (Issue 9) StandardBRCGS Storage and Distribution Issue 4 (3.1.3.1): reflecting shelf life and customer or legal requirements, never less than 1 year.
Storage and transport of time and temperature sensitive products (WHO model guidance, for reference)Temperature (and humidity) records for each storage environment; records in generalTemperature records: at least 1 year after the end of the stored product's shelf life, or as long as national legislation requires. Records generally: the minimum under local legislation, but not less than 3 years.GuidanceWHO Technical Report Series 961 (2011), Annex 9, clauses 9.3.1 and 9.2.3WHO TRS 961, Annex 9: Model guidance for the storage and transport of time- and temperature-sensitive pharmaceutical productsInternational model guidance, not UK or Irish law. National rules take precedence.
Medical device manufacturers and importers (EU MDR), IrelandManufacturers: technical documentation, EU declaration of conformity and any relevant certificates, with amendments and supplements. Importers: a copy of the EU declaration of conformity and any relevant certificates.At least 10 years; at least 15 years for implantable devicesLawRegulation (EU) 2017/745, Article 10(8) (manufacturers: 'at least 10 years after the last device covered by the EU declaration of conformity has been placed on the market. In the case of implantable devices, the period shall be at least 15 years') and Article 13(9) (importers keep copies 'for the period referred to in Article 10(8)')Regulation (EU) 2017/745 on medical devices (MDR), Articles 10(8) and 13(9)This covers device documentation, not temperature records as such. It is included as an example of a longer period that sits beside GDP. GB and NI device rules were not checked here.

These are differences between regimes, not contradictions: each period applies to its own records. Sources checked 19 October 2026.

Ireland: quality agreements often ask for longer

In Ireland, many pharmaceutical companies require their distributors to keep temperature records for 15 years through quality agreements; GDP itself requires at least 5. If a customer asks for 15, that's a contractual requirement: build it into your retention settings. This is what we see in practice, not a regulatory figure, so it is a difference between contract and law rather than a contradiction. When a customer or auditor quotes 15 years, ask which clause they rely on: usually it is the quality or technical agreement, sometimes a group data integrity policy. Other laws can also set longer periods for other records; for example, the EU MDR keeps device technical documentation for at least 10 years after the last device is placed on the market (15 for implantable devices).

Worked example: a hospital fridge that holds blood

Illustrative example. A hospital transfusion laboratory has a blood bank fridge that also holds a few ward medicines. The medicines alone would need at least 5 years under SPS and EU GDP, but the blood brings in RCPath's 15 years for the fridge charts and the BSQR's 30 years for traceability data. The BSQR's 30 years covers traceability data, not temperature charts, so the site sets 15 years for this fridge's temperature records and calibration certificates, keeps its transfusion traceability data for 30 years in the laboratory system, and sets the monitoring system to keep the temperature data, with its audit trail, for 15 years. It also records this decision in its SOP and checks once a year that the archived data still opens.

Sematics, the platform JTF installs and supports, keeps your records, unaltered, for as long as you're with us, and hands you the complete set, audit trail included, in readable formats if you ever leave.

Records and data integrity

Questions, answered from the source

Last checked 19 October 2026. Each answer links to its source and says whether it is law, formal guidance or interpretation.

How must retention and reference samples be stored, and for how long?Guidance

The revised Annex 19 (applicable from 24 September 2026) says reference and retention samples of each finished product batch should be kept for at least one year after the expiry date (3.1), stored in line with the guidance on declaration of storage conditions (5.1) and the marketing authorisation, for example refrigerated where relevant (5.2).

What's agreed, and what isn't

MHRA's 2015 GMP deficiency examples include a retention sample store with no temperature logger nearby, and a site with no storage facilities, monitoring devices or procedures for its reference and retention samples. The sample store is often the least watched room on site.

Status
Formal guidance
Applies to
Manufacturers, Laboratories
Last checked
19 October 2026
Can a dental practice keep its GlucaGen HypoKit in the emergency kit instead of the fridge?Interpretation

Yes. CQC dental mythbuster 4 says GlucaGen HypoKit should be stored at 2-8°C and not frozen, but it can be kept outside the fridge at no more than 25°C for 18 months, provided the manufacturer's expiry date is not exceeded, in its original package to protect it from light. The practice must be able to show either when the product came out of the fridge (for example a label with that date or a revised expiry) or that it is still safe, for example a purchase invoice showing that 18 months have not passed since delivery.

What's agreed, and what isn't

The 25°C limit means the room or kit location itself needs to stay below 25°C, and the mythbuster does not say how that should be evidenced. The 18-month figure is product-specific to GlucaGen HypoKit.

Status
Interpretation
Applies to
Dental
Last checked
18 October 2026
How long must vaccine fridge temperature records be kept?Guidance

The Green Book says records should be readily accessible, kept for at least one year, and cover the full storage history of anything in the fridge. Because vaccine shelf lives can be four years or more, it says keeping records for five years will generally cover the full storage history.

What's agreed, and what isn't

Strictly, the minimum is one year plus whatever covers the stock's storage history. Five years is a practical rule of thumb, not a fixed requirement. Northern Ireland simply says five years for daily records and cool box records. Scotland mirrors the Green Book wording (life of the stock, at least one year, five years advised).

Status
Formal guidance
Applies to
GP practices, Community pharmacy, Care homes
Last checked
18 October 2026
What maintenance records should we keep for a vaccine fridge?Guidance

The Green Book says to keep records of regular servicing, defrosting and cleaning, calibration and electrical (PAT) testing, on a log sheet kept with the fridge. Each fridge should have a unique identifier such as a serial number. It should be serviced at least yearly, sited in a ventilated place away from heat, and kept free of ice build-up.

What's agreed, and what isn't

If the fridge must be defrosted, the Green Book says to move vaccines to another fridge or a validated cool box for as short a time as possible. Scotland notes most incidents involve fridges over five years old and suggests planned replacement.

Status
Formal guidance
Applies to
GP practices, Community pharmacy
Last checked
18 October 2026
What routine checks of vaccine stock and fridge records does guidance expect?Guidance

The 2014 protocol asks for weekly checks of fridge contents (rotation and expiry), monthly stock audits, and three-monthly audit records that can be shared with the local screening and immunisation team, plus a quarterly vaccine management review. The Green Book asks for stock records to be checked and updated at least monthly and includes a monthly review line on its sample temperature chart.

What's agreed, and what isn't

The 2014 PHE protocol has not been revised since; UKHSA still hosts it but it refers to NHS England screening and immunisation teams and older NPSA/HPA links. Scotland asks for an annual site audit; Northern Ireland asks for weekly data logger reviews and an annual audit.

Status
Formal guidance
Applies to
GP practices, Community pharmacy
Last checked
18 October 2026
CQC found gaps in our vaccine fridge temperature records. Do we have to destroy the vaccines?Interpretation

No. PHE's Q&A says gaps or missing records do not on their own mean vaccines were stored badly. Do not dispose of vaccines or equipment; contact the local screening and immunisation team. Evidence such as stock inventories, delivery dates, stock turnover and the fridge's age, condition and service history can give reassurance.

What's agreed, and what isn't

This is a 2014 Q&A that refers to the withdrawn ImmForm helpsheet 18, but it matches the current 2022 UKHSA incident guidance (section 7.2: do not discard anything until discussed with the SIT/HPT). If there is real doubt, label the vaccines 'do not use' and quarantine the fridge while a risk assessment is done.

Status
Interpretation
Applies to
GP practices
Last checked
18 October 2026
Do we have to record wasted vaccines on ImmForm?Interpretation

Yes for centrally supplied stock. UKHSA asks that vaccines judged compromised are destroyed under local waste policy and recorded through the ImmForm Stock Incident page; expired stock should also be recorded there. UKHSA uses these reports to track wastage: £6.4 million reported in 2025, of which 84% was avoidable.

What's agreed, and what isn't

UKHSA describes ImmForm reporting as voluntary in its 2026 wastage article, and suspects under-reporting. This is separate from local incident reporting and from reporting to the SIT. The old ImmForm helpsheet 18 on fridge failures has been withdrawn.

Source
Vaccine update: issue 369, February 2026 UKHSA, 11 March 2026
Status
Interpretation
Applies to
GP practices, Community pharmacy
Last checked
18 October 2026
How long can a unit of red cells be out of the blood fridge and still go back into stock?Guidance

Where possible, keep red cells out of 2-6°C controlled storage for under 30 minutes. If 30 minutes is exceeded, the unit should not go back into the issue fridge but should be returned to the transfusion laboratory or quarantined remotely with electronic tracking. Up to 60 minutes is acceptable if the unit is then quarantined in a secure fridge for at least 6 hours before reissue, and a unit should not have more than three 30-60 minute episodes. Transfusion should be completed within 4 hours of leaving controlled temperature.

What's agreed, and what isn't

Hospitals need tracking (often electronic blood-tracking systems) to count episodes per unit; without it the three-occasion limit cannot be shown.

Status
Formal guidance
Applies to
Blood, Hospitals and wards
Last checked
18 October 2026
What do we need to write down after a temperature excursion?Guidance

SPS says to document every excursion locally, however minor, so patterns can be spotted. The minimum record is: who you are, when the excursion happened and how far out of range it went, the medicines and batch numbers affected, what advice was sought and from whom, and the actions taken to stop it happening again. The event should also go through your adverse incident procedure.

What's agreed, and what isn't

SPS also asks you to note whether a product has had a previous excursion. This matters because stability allowances are not unlimited and multiple excursions add up. SW ICARS makes the same point for vaccines.

Source
Managing temperature excursions NHS SPS, 9 July 2024
Status
Formal guidance
Applies to
All settings
Last checked
18 October 2026
In a GLP lab, do refrigerators with data capture need a user requirement specification and mapping?Interpretation

In its 2016 laboratories symposium Q&A, the MHRA said any equipment using electronic data capture is within scope of OECD GLP guidance document No. 17 on computerised systems. The level of validation depends on complexity. Where a computerised system captures or monitors data, an assessment of fitness for use is expected, including a specification covering temperature range and alarms. The refrigerator should be fit for purpose and able to deliver the required performance, which 'may require equipment validation such as temperature mapping'.

What's agreed, and what isn't

The answer is framed for GLP test facilities. Mapping is described as possibly required, not mandatory, so the extent should be risk-based and documented.

Status
Interpretation
Applies to
Laboratories
Last checked
18 October 2026
Do we have to trend our temperature and environmental monitoring data, or just check it?Guidance

Chapter 6, 6.9 says some kinds of data, including environmental controls, "should be recorded in a manner permitting trend evaluation", and out-of-trend data investigated. QC must also have environmental monitoring data and calibration records to hand (6.7). Chapter 4 requires procedures and records for environmental monitoring and calibration (4.29) and logbooks for critical equipment (4.31).

What's agreed, and what isn't

A daily tick that a reading was in range does not allow trend evaluation. A continuous electronic record does, provided someone reviews it. Annex 1 goes further for sterile sites (alarm events evaluated for trends, 5.2).

Status
Formal guidance
Applies to
Manufacturers, Laboratories
Last checked
19 October 2026
In Northern Ireland, how often should GP practices check vaccine fridges and how long should records be kept?Guidance

PHA guidance says vaccine fridge temperatures should be checked and documented at least daily and preferably twice daily (start and end of the working day), recording date, time, and minimum, maximum and current temperatures, then resetting the thermometer, including after periods of high activity. Fridges with an audible alarm are recommended and the alarm should be tested regularly. Daily temperature records should be kept for five years. Each fridge should have a unique identifier recorded on its temperature and servicing records, and satellite fridges must not be overlooked.

What's agreed, and what isn't

Twice-daily checking and five-year retention are stricter than the CQC (England) mythbuster's once per working day. This is NI regional guidance from the PHA, not an inspectorate (RQIA) standard.

Source
Guidance on vaccine handling and storage in GP practices Public Health Agency (NI), 30 September 2024
Status
Formal guidance
Applies to
GP practices
Last checked
18 October 2026
What should we look for when choosing a continuous temperature monitoring system?Guidance

WHO Supplement 6 recommends starting with a user requirements specification and choosing between an in-house hosted system and supplier-hosted SaaS. Wireless sensors should buffer data during network or power outages, sensors should be calibrated annually with a plan that avoids disrupting monitoring, and the system should have audit trails, secured data, different user levels, SOPs covering installation, use, backup and decommissioning, and full IQ/OQ/PQ. Alarm limits should be set only by authorised users, with automatic alerts by email or SMS and alert schedules for weekends and holidays. Round-the-clock technical support should be in the contract.

What's agreed, and what isn't

The supplement says that under SaaS the supplier ensures system maintenance and qualification. MHRA does not accept vendor validation in isolation from your intended use, so the user still has to verify. WHO notes that a simple 30-day temperature recorder may be enough for small peripheral stores.

Status
Formal guidance
Applies to
Wholesalers (WDA), Hospitals and wards, Laboratories, Community pharmacy
Last checked
18 October 2026
How often should we read the fridge thermometer, and what do we do if it's out of range?Interpretation

The MHRA blog says refrigerated stock should be monitored daily with a calibrated min/max device that has a calibration certificate. Records should show any deviation and the corrective action taken. As best practice, take a further reading later the same day to show the deviation was transient and the temperature is back in range. The RP should be informed of any deviation and should routinely review and sign off the records.

What's agreed, and what isn't

'Daily' is the minimum the blog describes; continuous electronic monitoring is now common and EU GDP expects alarms. The blog also expects records to be free from alterations and a trained deputy to cover absences. The blog does not say how to judge whether stock is still usable; that needs RP assessment and manufacturer stability data.

Status
Interpretation
Applies to
Wholesalers (WDA)
Last checked
18 October 2026
How should a care home record medicines fridge temperatures to satisfy CQC?Interpretation

Record the fridge temperature at least once a day, noting the minimum, maximum and current temperatures from a min/max thermometer (CQC prefers an inbuilt digital thermometer), and reset it after each reading so the record shows variation between readings. Staff must understand how to read and reset it, know what to do if the fridge is out of range, and record actions taken, including asking a pharmacist whether the medicines are still safe to use. Make sure the probe cable does not interfere with the door seal.

What's agreed, and what isn't

CQC asks for 'at least once a day', which in a 24/7 care home means every day including weekends. That is stricter than the GP 'each working day' wording. CQC does not specify calibration or record retention for care homes.

Status
Interpretation
Applies to
Care homes
Last checked
18 October 2026
If we have a data logger in the vaccine fridge, do we still need to do daily manual readings?Interpretation

Yes. CQC says that even with a data logger you must still read and record the integral thermometer's minimum, maximum and current temperatures and reset the min/max thermometer each working day. It warns that checking a logger only weekly or monthly risks missing cold chain breaches and using vaccines that were stored out of range. It also asks whether an alarm could clear itself once the temperature returns to range without anyone investigating. CQC sees loggers as useful for detailed information after a failure such as a power cut.

What's agreed, and what isn't

The mythbuster does not say whether a continuously monitored, alarmed remote system with an audit trail could replace manual readings. As written, it expects the daily manual record regardless, so check with your ICB before dropping paper records. VMD and NI PHA guidance take the same 'logger plus daily check' position.

Status
Interpretation
Applies to
GP practices
Last checked
18 October 2026
What thermometer or data logger set-up does the VMD expect in a vet practice, and how often should it be checked?Interpretation

As a minimum the VMD recommends a digital thermometer showing current, minimum and maximum temperatures. For all refrigerated storage, including vehicle fridges and vaccine chillers, record min and max daily and reset after each reading. Ambient areas should be recorded daily, though this can drop to weekly where temperatures are consistently stable. Data loggers must be set to flag excursions; for fridges, download and review them at least weekly, with logging roughly every 15-30 minutes. They do not replace daily checks that the device is working and no alert has fired.

What's agreed, and what isn't

The VMD notes that min/max readings cannot show how long an excursion lasted, which is an argument for loggers. It suggests insulating cabled probes to mimic product temperature and choosing a calibrated system, but gives no calibration interval. Using a third-party monitoring service does not transfer responsibility for responding to alerts.

Status
Interpretation
Applies to
Veterinary
Last checked
18 October 2026
How long must batch records for an investigational product be kept?Law

At least five years after the completion or formal discontinuation of the last clinical trial in which the batch was used. For Great Britain the period comes from Commission Directive 2003/94/EC, Article 9(1), which the UK clinical trials regulations still reference; Article 9(2) adds that electronic systems must be validated, data backed up and audit trails maintained. For Northern Ireland, Delegated Regulation 2017/1569, Article 8(3) sets the same five years.

What's agreed, and what isn't

Sample retention for IMPs is two years after the last trial ends (2017/1569, Article 11). Keep the storage and shipping temperature records that support a batch's certification for at least as long as its batch documentation.

Source
Commission Directive 2003/94/EC, Article 9 (Documentation) European Commission, 8 October 2003
Status
Law
Applies to
Manufacturers
Last checked
19 October 2026
Can all our staff use one shared login for the temperature monitoring software?Guidance

No. MHRA says shared logins or generic user access should not be used for systems that generate, amend or store GXP data, and where a system supports individual accounts this must be used, even if it means buying extra licences. Companies must be able to show the access level given to each person, including historical access. System administrator access should be limited to as few people as possible, with unique credentials, and should not be given to people with a direct interest in the data (those generating, reviewing or approving it).

What's agreed, and what isn't

Where a system genuinely supports only one login, MHRA accepts equivalent control via third-party software or a version-controlled paper method, justified and documented, with increased data review. PIC/S PI 041-1 goes further and says shared passwords "even for reasons of financial savings, should be prohibited". Small sites may struggle with the segregation of admin rights, so document the rationale.

Status
Formal guidance
Applies to
All settings
Last checked
18 October 2026
Can our QA lead or the person who reviews the data also be the administrator of the monitoring system?Guidance

The MHRA's GxP data integrity guidance says shared logins or generic user access should not be used for systems generating, amending or storing GxP data, and that system administrator rights "should not be assigned to individuals with a direct interest in the data (data generation, data review or approval)" (6.16). Audit trails identified as required should be switched on and users should not be able to amend or switch them off (6.13).

What's agreed, and what isn't

Administrator access should be kept to the fewest people possible for the size of the organisation, each with unique credentials so their actions show in the audit trail. In a cloud system the supplier usually holds platform administration; your own administrator roles still need the same separation.

Status
Formal guidance
Applies to
Manufacturers, Laboratories, Wholesalers (WDA)
Last checked
19 October 2026
Can we delete or ignore a temperature reading that we think was a faulty sensor spike?Guidance

Data may only be excluded where valid scientific justification shows it is not representative of what was measured. The justification must be documented and considered during data review. All data, even if excluded, should be kept with the original data set in a form that lets someone confirm the decision to exclude was valid.

What's agreed, and what isn't

In practice a suspected sensor fault should be handled as a deviation: investigate it, check the sensor (and calibration), and record the conclusion against the reading rather than removing it. MHRA does not give temperature-specific examples here.

Status
Formal guidance
Applies to
All settings
Last checked
18 October 2026
Can we scan our paper temperature record sheets and then destroy the paper originals?Guidance

Yes, if the scan is a verified true copy. MHRA allows true copies to be kept in place of the original (for example a scan of a paper record) where a documented system verifies and records the integrity of the copy, by a dated signature or a validated process. Paper data may be kept using a validated scanning process, provided a documented process ensures the outcome is a true copy. Organisations should consider any risk of destroying the originals.

What's agreed, and what isn't

The guidance does not say how to verify each scan. A 'validated scanning process' is defined as one with added controls so each page need not be checked by a person. Without that, someone needs to check and sign each copy. Make sure the scan is legible for the whole retention period and keeps signatures, dates and corrections visible.

Status
Formal guidance
Applies to
All settings
Last checked
18 October 2026
Does a simple digital fridge thermometer need validating as a computerised system, or is calibration enough?Guidance

MHRA says simple electronic devices with no configurable software and no electronic data retention, and it names thermometers as an example, "may only require calibration". More complex systems need validation for their intended purpose, and the effort grows with configurability, user intervention and data lifecycle risk. Where basic equipment does store data but overwrites it after a set volume, the data should be reviewed periodically, reconciled with paper records where needed, and extracted electronically where the device supports it.

What's agreed, and what isn't

A data logger with download software, configurable alarms or a cloud dashboard is not a 'simple' device in this sense, and validation expectations apply. MHRA also warns not to overlook apparently simple systems where data could be manipulated or repeated without detection.

Status
Formal guidance
Applies to
All settings
Last checked
18 October 2026
Does an electronic temperature monitoring system need validating under HPRA GDP guidance?Guidance

The HPRA GDP guide says a computerised system relating to GDP processes should be shown, through validation or verification, to achieve results accurately, consistently and reproducibly before use, with the level of validation set by risk (complexity, bespoke vs off-the-shelf, customisation). It points to EU GMP Annex 11 for further information, and equipment qualification protocols should be approved by the RP.

What's agreed, and what isn't

The guide does not single out temperature monitoring systems or give data integrity (ALCOA+) detail; no HPRA GDP-specific data integrity guide was found. IA-G0011 adds that records must be retrievable, allow excursions to be identified, and that reviews must be recorded.

Status
Formal guidance
Applies to
Wholesalers (WDA)
Last checked
18 October 2026
Does our electronic temperature monitoring system need an audit trail, and do we have to review it?Guidance

Where a computerised system captures, stores or reports raw data, MHRA expects it to keep audit trails showing all changes and deletions (who, what, when and why) while keeping the original data. Audit trails identified by risk assessment as required should be switched on, and users should not be able to amend or turn them off. Routine data review should include a documented audit trail review where risk assessment requires it, and this can be done by a validated exception report. It need not cover every log-on or keystroke.

What's agreed, and what isn't

For legacy systems without an audit trail, alternative controls such as an SOP and log books are allowed if shown to be effective, but MHRA expects demonstrated progress towards a compliant system and may cite a deficiency if remediation is not timely. The guidance does not set a review frequency; that comes from your own risk assessment. For a monitoring system, relevant entries include alarm limit changes, sensor reassignment, acknowledgements and any edits to readings.

Status
Formal guidance
Applies to
All settings
Last checked
18 October 2026
Does our electronic temperature monitoring system need to be validated, and how long do we keep the data?Guidance

EU GDP 3.3.1 says a computerised system should be shown, through validation or verification, to achieve the desired results accurately, consistently and reproducibly before it is used. Data should be protected against accidental or unauthorised change, checked periodically for accessibility, and backed up regularly. Backups should be kept for the period in national legislation but at least five years, at a separate and secure location.

What's agreed, and what isn't

The guideline also expects a written description of the system and defined procedures for system failure, including data restoration. It does not prescribe audit trail features; the MHRA GxP data integrity guidance gives more detail. Cloud-hosted systems still need the separate, secure backup location to be demonstrable.

Status
Formal guidance
Applies to
Wholesalers (WDA)
Last checked
18 October 2026
How long do we have to keep temperature records?Guidance

EU GDP chapter 4 says documents should be kept for the period stated in national legislation but at least five years, and records should be made at the time each operation is undertaken. Chapter 3.3.1 applies the same five-year minimum to backups of computerised system data.

What's agreed, and what isn't

Five years is a minimum; manufacturer, contract or other legal requirements may require longer. Records must be contemporaneous, so filling gaps afterwards is a data integrity failure (the 2016 MHRA deficiency data cites an RP who added fabricated readings).

Status
Formal guidance
Applies to
Wholesalers (WDA)
Last checked
18 October 2026
How long do we need to keep fridge and room temperature records?Guidance

SPS's pharmacy records retention guidance, republished in September 2026, gives at least 5 years for medicine storage temperature records. This covers fridge and ambient readings, excursions, monitoring-system downloads and calibration records, and is based on GDP record-retention advice. For vaccines, it also gives at least 5 years for temperature logs and 11 years for refrigerator maintenance logs.

What's agreed, and what isn't

The Green Book (chapter 3), which SPS quotes, sets a floor of at least 1 year but says records should cover the full storage history of products held. Because vaccine shelf lives can be 4 years or more, SPS says 5 years generally covers it. The RPS guidance and NICE SC1 set no retention period. Care homes and GP practices often keep records for 1-2 years, which is shorter than the SPS figure.

Status
Formal guidance
Applies to
All settings
Last checked
18 October 2026
How long must a manufacturer keep storage temperature records?Guidance

Chapter 4 does not name temperature records. Batch documentation must be kept for one year after batch expiry or at least five years after QP certification, whichever is longer (4.11). Critical documentation, including raw data relating to validation or stability that supports the marketing authorisation, must be kept while the authorisation is in force (4.12). Records must stay intact and readable for the whole period (4.1, 4.10).

What's agreed, and what isn't

Storage records that support a batch's release fall in practice with its batch documentation, and chamber records behind stability data with the MA-supporting raw data; this is a reading of 4.11 and 4.12, not their words. Longer periods may apply to products such as ATMPs (4.11). For investigational products the batch documentation period is five years after the last trial using the batch ends.

Status
Formal guidance
Applies to
Manufacturers, Laboratories
Last checked
19 October 2026
How long must a tissue establishment keep temperature monitoring records for tissues and cells used in patient treatment?Guidance

The HTA Guide to Quality and Safety Assurance says raw data such as temperature monitoring printouts must be kept for at least 10 years in their original format, and records critical to quality and safety for at least 10 years after expiry, clinical use or disposal. All critical equipment must have set operational limits, such as temperature, and be monitored, with calibration against a traceable standard where available and procedures for malfunction or failure. Transport conditions, including temperature and time limits, must be specified and validated, and shipping containers validated for their purpose.

What's agreed, and what isn't

After 10 years the guide allows raw data to be stored in another format. For electronic monitoring systems, 'original format' means the native data with its audit trail, which needs planning for system migrations.

Status
Formal guidance
Applies to
Tissue and mortuary
Last checked
18 October 2026
How long must an API manufacturer keep records, including storage records?Guidance

Part II, 6.13 says records should be kept for at least one year after the batch's expiry date. For APIs with retest dates, records should be kept for at least three years after the batch is completely distributed.

What's agreed, and what isn't

Part II also requires controls to prevent omissions in data, for example "system turned off and data not captured" (5.43), and a back-up system where a breakdown would mean permanent loss of records (5.48). Both apply directly to a monitoring system.

Status
Formal guidance
Applies to
Manufacturers
Last checked
19 October 2026
How long should we keep temperature records?Guidance

WHO's model guidance gives two figures. Clause 9.3.1 says temperature records should be kept for at least one year after the end of the shelf life of the stored product, or as long as national law requires. Clause 9.2.3 says records generally should be kept for the minimum period under local law, but not less than three years. WHO Supplement 15 says downloaded transport logger data should in most cases be kept at least three years in a retrievable, non-volatile format.

What's agreed, and what isn't

The WHO document is inconsistent with itself, and national law takes precedence. In the UK, check the retention period required by your licence, GDP guidance or sector guidance. EMA's GMP Q&A on Annex 11 confirms electronic records have the same retention requirements as paper.

Status
Formal guidance
Applies to
Wholesalers (WDA), Hospitals and wards, Laboratories, Transport
Last checked
18 October 2026
Once our GMP temperature monitoring system is validated, do we need to do anything else to keep it validated?Guidance

Yes. EU GMP Annex 11 requires computerised systems to be periodically evaluated to confirm they remain valid, looking at things like deviations, incidents, upgrades, performance, security and validation status. Backups should be made regularly and restoration checked, stored data checked for accessibility and readability, and audit trails reviewed regularly. Changes, including configuration changes, must go through a defined change procedure, and it should be possible to get clear printed copies of stored data.

What's agreed, and what isn't

Annex 11 sets no fixed interval. The EMA Q&A says the period for re-evaluation should be based on the system's criticality. Annex 11 is EU GMP guidance that continues to be applied in UK GMP. A revised Annex 11 has been in draft, so check for updates.

Source
EudraLex Volume 4, Annex 11: Computerised Systems European Commission, January 2011
Status
Formal guidance
Applies to
Manufacturers, Laboratories
Last checked
18 October 2026
Our logger software backs up to a server every night. Does that count as archiving our temperature records?Guidance

No. MHRA distinguishes backup (a copy of current data for disaster recovery) from archive (long-term protected retention), and says backups "do not replace the need for the long term, retention of data". Backup and recovery should be validated and periodically tested, and each backup checked to confirm it worked. Archived records must be protected from undetected change or deletion and stay retrievable and readable, with metadata, throughout the retention period. Before a system is decommissioned, its records should be archived under a documented strategy, for example by keeping the software in a virtual environment or migrating to a format that keeps as much of the true copy as possible.

What's agreed, and what isn't

The MHRA guidance does not set a retention period for temperature data; it defers to the legislative requirements for each GXP area. Plan for what happens to historical logger data when you change monitoring supplier.

Status
Formal guidance
Applies to
All settings
Last checked
18 October 2026
Our monitoring gateway was offline for a weekend and readings are missing. Is that a GMP problem?Guidance

It can be. Part II says computerised systems should have controls to prevent omissions in data, such as a system turned off and data not captured, and a record of any data change (5.43); where a breakdown could permanently lose records, a back-up system should be provided (5.48). Annex 11 asks that data stay accessible throughout the retention period and backups be checked for restoration (7.1, 7.2).

What's agreed, and what isn't

A gap is a deviation to investigate: when it started, whether the sensors buffered readings and backfilled them, and whether the product was at risk. A monitoring system that alarms when a sensor or gateway goes offline turns a silent gap into a prompt to act.

Status
Formal guidance
Applies to
Manufacturers, Laboratories
Last checked
19 October 2026
Our temperature monitoring system is cloud-hosted by the supplier. What does the MHRA expect us to have in place?Guidance

MHRA says you should understand the service provided, who owns the data, and how it is retrieved, retained and secured, including where the data is physically held and which laws apply there. Responsibilities should be set out in a technical agreement or contract that guarantees timely access to data, metadata and audit trails for you and for the authorities on request, and covers archiving and readability for the whole retention period. Arrangements must exist to restore the system to its validated state, business continuity should be in the contract and tested, and the need to audit the provider should be based on risk.

What's agreed, and what isn't

WHO Technical Supplement 6 describes software-as-a-service as an arrangement where the supplier "ensures the system maintenance and qualification". MHRA is clear that the regulated company keeps responsibility for the data and for validation for its intended use. Check what happens to your records if the contract ends or the supplier fails.

Status
Formal guidance
Applies to
All settings
Last checked
18 October 2026
Our temperature monitoring system stores readings electronically. Can we just print and file the weekly reports instead of keeping the electronic data?Guidance

MHRA says raw data is the first capture of information, and where it was captured electronically in a dynamic state "paper copies cannot be considered as 'raw data'". Where a system can store data dynamically, it is not appropriate to keep static (printed or manual) records in preference to the electronic data. Retaining only a static record is possible only where it can be justified that it keeps the integrity of the original, including verified copies of raw data, metadata, audit trails and configuration settings, plus a documented check that the printouts are accurate. MHRA notes this approach is "likely to be demanding in its administration".

What's agreed, and what isn't

WHO Technical Supplement 15 (transport monitoring) lists printed hard copies as an acceptable retention format for downloaded logger data, so the WHO position is looser than MHRA's. For UK-regulated activity the MHRA expectation is the one inspectors apply. Printouts remain useful for review and signing, as long as the electronic record is kept as the original.

Status
Formal guidance
Applies to
All settings
Last checked
18 October 2026
The monitoring system supplier gave us a validation pack. Is that enough to show the system is validated?Guidance

Not on its own. MHRA says computerised systems must be validated for their intended purpose, and "the acceptance of vendor-supplied validation data in isolation" from your configuration and intended use is not acceptable. Vendor testing done away from your process and IT setup is likely to be functional verification only, and may not meet performance qualification requirements. Vendor documents can feed into your validation, but you still need to show the system works as configured at your site.

What's agreed, and what isn't

How much site testing is needed depends on risk and complexity (MHRA section 4.3). WHO Supplement 6 expects installed monitoring systems to be qualified through IQ, OQ and PQ after a commissioning period. For SaaS systems the supplier may do much of the work, but the user still has to confirm the system meets its own requirements.

Status
Formal guidance
Applies to
Wholesalers (WDA), Manufacturers, Hospitals and wards, Laboratories, Blood, Tissue and mortuary
Last checked
18 October 2026
We are a small operation. How much data integrity control does the MHRA expect for our temperature records?Guidance

MHRA describes its guidance as a minimum expectation but asks for a risk-based approach where effort is "commensurate with the risk and impact" of a data integrity failure. Organisations are not expected to check data forensically as a routine, and a data integrity risk assessment is suggested to map processes and controls. Switching from electronic to paper, or back, does not remove the need for appropriate controls. The guidance encourages talking to the MHRA where your approach differs from what it describes.

What's agreed, and what isn't

The guidance does not cover medical devices. It is written for all MHRA-regulated GXP areas, so the examples are not always temperature-specific; you have to apply the principles to your own monitoring set-up. The MHRA Inspectorate blog (9 March 2018) confirmed the guide covers GDP as well as GMP and supersedes the March 2015 GMP-only guide.

Status
Formal guidance
Applies to
All settings
Last checked
18 October 2026
We read the fridge display and write the temperature on a paper chart. Is that acceptable from a data integrity point of view?Guidance

Yes. A paper record of a manual observation can be the original record, but MHRA expects the risk to be assessed, and depending on how critical the data is, a second check or another way of capturing it may be justified. Records should be available where the work happens so people do not jot readings informally and copy them later, and blank forms should be controlled (for example numbered and reconciled). Where readings are transcribed from one record to another, the transcription should be checked by a second person or a validated system.

What's agreed, and what isn't

WHO TRS 1033 Annex 4 also gives writing a logger reading on a room log sheet as an example of an original record. If the device itself stores the data electronically, MHRA's view is that the stored electronic data is the raw data, and the paper chart is a secondary record.

Status
Formal guidance
Applies to
All settings
Last checked
18 October 2026
We use a mix of paper logs and an electronic logger. Is a hybrid system acceptable?Guidance

Hybrid systems are allowed, but MHRA says it should be clearly documented what makes up the whole data set, and all records in that set should be reviewed and kept. MHRA also expects more data review for hybrid systems, because they are vulnerable to changes that cannot be attributed to anyone. WHO's data integrity guideline says where hybrid systems are used, even for an interim period, the risks should be assessed with documented mitigating controls, and replacing them "should be a priority with a documented CAPA plan".

What's agreed, and what isn't

The WHO view (replace hybrids as a priority) is firmer than MHRA's (control and review them). For small sites with a fridge logger plus a daily paper check, documenting which record is the original and how the two are linked is the practical minimum.

Status
Formal guidance
Applies to
All settings
Last checked
18 October 2026
What should our documented review of temperature records actually include?Guidance

MHRA expects a procedure for reviewing and approving data, including a risk-based review of relevant metadata and audit trails. The review should be documented with a positive statement of whether or not issues were found, the date of the review and the reviewer's signature. A procedure should say what happens when an error or omission is found, with corrections that keep the original visible and traceable. Where you rely on summary reports from another organisation, such as a courier's or 3PL's temperature summary, MHRA says summaries are limited and you should assess the provider's data integrity controls first.

What's agreed, and what isn't

The guidance does not set how often reviews happen. WHO TRS 961 Annex 9 suggests reviewing temperature records monthly, with checks at least twice daily. UK sector guidance sets its own frequencies, so check those for your setting.

Status
Formal guidance
Applies to
All settings
Last checked
18 October 2026
Can we keep temperature records in a shared spreadsheet or report alarms over WhatsApp?Interpretation

The MHRA data integrity blog for GDP says multi-user spreadsheets can make it impossible to show who made an entry, when, or whether it was overwritten. It suggests restricting spreadsheets to non-critical data, or locking and version-controlling them. It says a wholesaler's proposal to use WhatsApp for quality-critical event reporting was refused because it could not meet data integrity requirements. It also gives an example where an inspector could not trust ambient min/max records because the thermometer was uncalibrated and staff did not know how to use it.

What's agreed, and what isn't

Computerised systems are encouraged if appropriately qualified and controlled. The MHRA GxP Data Integrity guidance (2018) gives the detailed ALCOA expectations. The MHRA's 2016 GDP deficiency data shows what happens otherwise: gaps in temperature records that an RP admitted filling with fabricated readings, entries in pencil, unsigned records and unapproved corrections, all cited as major deficiencies. Missed readings should be recorded as missed, never back-filled. The blog also says paper temperature records transcribed into an electronic system should be checked by a person of appropriate seniority.

Source
I don't believe it! MHRA, 25 May 2018
Status
Interpretation
Applies to
Wholesalers (WDA)
Last checked
18 October 2026
Is it acceptable to record or analyse temperature data in an Excel spreadsheet?Interpretation

EMA's GMP Q&A says spreadsheet templates should be checked for accuracy and version controlled, and formulas and files should be secured so they are not accidentally overwritten. Entry of the wrong data type should be blocked or flagged, and 'boundary checks' are encouraged. Formal validation is required at least for spreadsheets with custom code such as macros, and formulas should be verified. MHRA adds that electronic worksheets should be version controlled, and that data transfer into a worksheet should be validated and audit-trailed.

What's agreed, and what isn't

None of these sources says directly whether a spreadsheet can be the primary temperature record. Ordinary spreadsheets usually lack individual logins and audit trails, which MHRA expects for GXP data (sections 6.13 and 6.16). Using one as the master record therefore needs extra controls and justification. Using it to analyse data exported from a logger is easier to defend, as long as the original logger data is kept.

Status
Interpretation
Applies to
All settings
Last checked
18 October 2026
What data integrity failures in temperature records have MHRA GDP inspectors found at wholesalers?Interpretation

MHRA's 2016 GDP deficiency data includes a Responsible Person who admitted adding fabricated readings to fill gaps in temperature records. Other findings were records made in pencil, unsigned temperature records and unapproved corrections. Equipment findings included a warehouse thermometer not calibrated across its operating range, an ambient-area thermometer installed so readings were not taken from the calibrated probe, uncalibrated equipment used to monitor transport, and mapping that did not establish where to place thermometers.

What's agreed, and what isn't

The report is published to support self-inspection. It shows that inspectors treat gaps and fabricated or unattributed entries as serious. A missed reading should be recorded and explained as a gap, never filled in afterwards.

Status
Interpretation
Applies to
Wholesalers (WDA)
Last checked
18 October 2026
What GAMP category is a cloud temperature monitoring system?Interpretation

GAMP 5 (second edition, 2022) groups software as infrastructure, tools and IT services, standard system components, configured components, and custom applications and components (categories 1, 3, 4 and 5; category 2 has not been used since 2008). ISPE says categorisation applies to the components that make up a system and the categories should be viewed as a continuum. A configurable monitoring platform is usually assessed as configured (category 4), with any custom interfaces higher.

What's agreed, and what isn't

ISPE stresses that the software category is only one factor in a risk-based approach; the risk of each function to product and data matters more. GAMP 5 is industry guidance, not regulation, though inspectors know it well. Its second edition adds guidance on assessing cloud service providers and SaaS.

Status
Interpretation
Applies to
Manufacturers, Laboratories
Last checked
19 October 2026
Do we need to keep and review the temperature data from each delivery?Interpretation

The MHRA blog says best practice is calibrated temperature monitoring that gives data for the entire journey. The wholesaler should keep this data for the required five years. Distributors should review the data for each consignment and have procedures for corrective action when something goes wrong. Each journey should be treated as unique, with length, complexity and season considered when choosing packaging and transport.

What's agreed, and what isn't

Monitoring every shipment is described as best practice, not an absolute requirement. The 2020 MHRA transport blog adds that where shipments are not individually monitored, "a very robust validation is required with clearly defined limits of validation".

Status
Interpretation
Applies to
Wholesalers (WDA), Transport
Last checked
18 October 2026
Do the blood storage rules apply to a hospital blood bank, or only to blood establishments?Law

They apply to hospital blood banks too. Regulation 9 requires the person managing a hospital blood bank to keep a quality system that gives effect to the EDQM Good Practice Guidelines (reg 9(1)(b)), make sure storage, transport and distribution conditions meet Part 4 of the Schedule, and validate all Part 4 processes that apply to its activities. Documentation must be available for inspection, traceability data must be kept for at least 30 years, and serious adverse event records for at least 15 years.

What's agreed, and what isn't

Regulation 9 does not itself specify how storage must be monitored (continuous, alarms, mapping); that detail comes in through the EDQM Good Practice Guidelines the quality system must give effect to. MHRA inspectors apply those, GMP-style expectations and the Red Book in practice. The 30-year requirement covers traceability data, not necessarily raw temperature logs, so a local retention decision is needed for those. For blood bank refrigerator and freezer temperature charts, the RCPath and IBMS retention guidance (6th edition, October 2025, section 6.1.5) says 15 years; that is professional guidance, not law. The retention table on the records and data integrity page compares the periods.

Status
Law
Applies to
Blood, Hospitals and wards
Last checked
18 October 2026
How often does the Responsible Person have to review temperature records under HPRA guidance?Guidance

IA-G0011 repeatedly states that temperature records should be checked daily and independently reviewed (and, for ambient storage, approved) on a monthly basis by the Responsible Person, with a record of the review kept. Deviations outside limits should be notified promptly to the RP and investigations documented.

What's agreed, and what isn't

Where alarms notify staff audibly in real time and out of hours, the guide allows the frequency of checking monitoring records to be extended if the rationale is documented and justified. The EU GDP Guidelines do not specify a monthly RP review; this is HPRA-specific detail.

Status
Formal guidance
Applies to
Wholesalers (WDA)
Last checked
18 October 2026
Do the GPhC standards for registered pharmacies set a fridge temperature requirement?Interpretation

Not explicitly. Standard 4.3 says medicines and medical devices must be obtained from a reputable source, safe and fit for purpose, stored securely, safeguarded from unauthorised access, supplied safely and disposed of safely. There is no temperature wording. In practice, GPhC inspectors assess cold chain under 4.3. For example, a 2026 published improvement action plan recorded 4.3 as not met because a pharmacy recorded temperatures for only one of its two medicines fridges, and the unrecorded fridge showed an out-of-range maximum. The fixes included calibrated thermometers, daily recording for both fridges, and noting the reasons for anomalies.

What's agreed, and what isn't

The expectation of daily records and action on out-of-range readings comes from inspection practice and SPS/RPS guidance, not from the text of the standard. An individual action plan shows one inspector's finding and is not a GPhC policy statement.

Status
Interpretation
Applies to
Community pharmacy
Last checked
18 October 2026